05-NOVEMBER-2022 MARAAH HUSSEIN ABU-ELNADI 26
YEARS MALFUNCTIONING VPS.
Anamnesis
The patient was operated by me 26 years ago for
hydrocephalic and had dorso-lumbar syringomeylia
with paraplegia and anesthesia below D10. The
patient came to the clinic 01-November-2022
complaining of headache and neck pain for the
last 2 weeks. The patient was in other hospital
and CT-scan of the brain showed the shunt in
place. She is in Depakine Chrono twice daily for
long years.
On examination, the patient is in wheelchair
with complete paraplegia anesthesia below D10
with more spastic left lower limb and bobbing
nystagmus and left abducens paresis for 2 weeks.
No fever, nor meningism. The reservoir of the
shunt was functioning and the track of the
distal tube was palpable, but very tense at the
cranio-vertebral area. Plain abdomen showed the
distal part very short inside the abdominal
cavity.
The patient was sent for investigations and MRI
done 2-November-2022 showing the the
hydrocephalus and the shunt in place with
elements of Arnold Chiari malformation reaching
C2, but without compression. There is gross
deformity of the lower dorsal and upper lumbar
area with tethering of the spinal cord to this
level. It was advised to replace the VPS of
pediatric brand to adult brand, because when it
is tight the lumen of the tube becoming narrow
and malfunctioning, threatening rupture of the
device.
The patient was brought to the Emergency of
Shmaisani hospital 04-November-2002 with convulsions and episode
of blindness before convulsion attack. The
patient was given Epanutin I/V and Mannitol. The
seizure weaned off and vision returned.
Palpation of the distal part of the shunt was
hard to feel for what CXR done ruling out rupture of the
distal part of the shunt. The reservoir is not
functioning.
The old pediatric shunt was
replaced with new medium pressure Medtronic adult type of
VPS. The old one was surrounded by calcification
and the distal abdominal end short with its
slits in the abdominal wall. Several incision
were performed over the previous track, because
it was afraid that still having parts of the
tube, but it was pure calcification.
FOLLOW UP
Too early now.
Comments
The pediatric shunt usually must be
replaced at the age of 8-9 years. But here the patient came
after 28 years. When the shunt is severely stretched the
lumen becoming very narrow or the tube rupture. In both
cases it must be replaced with adult type of VPS to avoid
possible permanent blindness.
Skyra MRI with all clinical applications in the run since 28-Novemeber-2013.
Inomed Riechert-Mundinger System, with three point
fixation is the most accurate system in the market. The microdrive and
its sensor gives feed back about the localization.
Inomed MER system
Leica HM500
The World's first and the only Head mounted Microscope.
Freedom combined with Outstanding Vision, but very bad video recording and
documentation.
After long years TRUMPF TruSystem 7500 is running with in the neurosuite at
Shmaisani hospital starting from 23-March-2014
LooksCam II Xenosys in the run starting from 14-March-2021 with
SheerVision TTL x4 magnification.
Cios-Spin flat panel in the run.
Notice: Not all operative activities
can be recorded due to lack of time.
Notice: Head injuries and very urgent surgeries are also
escaped from the plan .